Tiivistelmä
The objective was to perform three EMC measurements, Site Voltage Standing Wave Ratio (SVSWR), Normalized Site Attenuation (NSA), and Field Uniformity measurements which are used for validating EMC test sites.
The work involved reading and interpreting the relevant standards in depth, performing the measurements, and presenting the results for use in the accreditation process. The measurement results showed compliance with accreditation requirements. Comprehensive site specific guides were developed for Verkotan to perform these measurements when necessary.
Additionally, the thesis explored other standards that use methodologies derived from EN 61000-4-3 and EN 55016-1-4, providing further context for their application in various EMC testing scenarios.
This thesis provided valuable support in establishing a solid foundation for EMC testing capabilities within the company.
The work involved reading and interpreting the relevant standards in depth, performing the measurements, and presenting the results for use in the accreditation process. The measurement results showed compliance with accreditation requirements. Comprehensive site specific guides were developed for Verkotan to perform these measurements when necessary.
Additionally, the thesis explored other standards that use methodologies derived from EN 61000-4-3 and EN 55016-1-4, providing further context for their application in various EMC testing scenarios.
This thesis provided valuable support in establishing a solid foundation for EMC testing capabilities within the company.